EasyManua.ls Logo

Keithley 4200A-SCS

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
H: Using a Micromanipulator 8860 Prober Model 4200A-SCS Parameter Analyzer
Reference Manual
H-12 4200A-901-01 Rev. C / February 2017
Load the wafer
Load the wafer:
1. Make sure that the vacuum is off.
2. Click the Load wafer button on the Tools panel of the pcNav window. The Load Wafer dialog box
appears. See the following two figures.
Figure 694: Load Wafer button
Figure 695: Load Wafer window
3. In the Load Wafer dialog box, click Load.
4. After the chuck moves to the front, place wafer on the chuck aligning the flat or notch in the
proper orientation.
5. Apply vacuum.
6. Click Center.
7. Click Done.

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals