-SCS Parameter Analyzer Reference Manual Section 6:
4200A-901-01 Rev. C / February 2017 6-129
SegARBConfig control parameters
The following tables summarize all the parameters in the SegARBConfig control. In the tables, each
parameter is shown mapped to its parameter target. As mentioned above, the Segment Arb
functionality is primarily in two LPT commands: seg_arb_sequence and seg_arb_waveform. In
addition, a few parameters are unique to the SegARBConfig control.
SegARBConfig parameter names mapped to LPT function seg_arb_sequence
In this table, the variable name is in the function seg_arb_sequence (on page 13-125
).
Parameter
name
Variable name in
function Type
Required in
SegARBConfig? Description
Double array Yes
Time duration for each
Integer array Yes
Solid State Relay control, per
segment
Trigger output state for each
segment
Integer
1
spot mean, or waveform
Double
1
Start point for measurement
window, for each segment;
from 0 to 1 (100%)
Double
1
Stop point for measurement
window, per segment; from 0
to 1 (100%)
Number of segments in a
sequence
If using a single value instead of an array, the SegARBConfig control automatically assigns the value to each
SegARBConfig parameter names mapped to SegARBConfig
In this table, the variable name is in the function SegARBConfig.
Parameter
name
Variable name in
function Type
Required in
SegARBConfig? Description
List of defined sequences in
SegARBConfig
Integer array No
Start segment array value in
Stop segment array value in
SegARBConfig
SegMeas parameter turns a measurement on or off for each segment and is independent of the SegMeasType
(based on implementation in user module).