-SCS Parameter Analyzer Reference Manual Section 8:
Keithley User Library Tool (KULT)
4200A-901-01 Rev. C / February 2017 8-61
5. Between the Place user library module call below and Place user library
module call above comments, call your user module as it is described in
<debugtaskname>.c. For the VSweep user module, you could enter the following code:
Double Imeas;
Double Vforce;
VSweep(0, 5, Imeas, 11, Vforce, 11)
You can also define a UTM in Clarius with your user module, as shown in Tutorial 2: Creating a user
module that returns data arrays (on page 8-28), Checking the VSweep user module (on page 8-32).
This generates the file <UTMname>.ktm at C:\s4200\Projects\<projectname>\tests, which
includes the necessary call to your user module. Open <UTMname>.ktm with a text editor, copy the
contents, and paste them into <debugtaskname>.c. Using this method ensures that the input
parameters for your user module fit the restrictions set by KULT.
6. In the Build menu, select Compile to compile the <debugtaskname>.c program.
Loading a debug task
To load a saved debug task, use the load_dt command-line utility.
c:\>load_dt <debugtaskname>
For example, to load the debug task called VSweepDebug, enter the following at the command line
prompt:
c:\>load_dt VSweepDebug
Figure 457: Example code generated by the create_dt utility