Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-29
WT
This command delays the start of a test sequence for time domain measurements.
Usage
WT AAA.AAA
AAA.AAA
Wait time in seconds: 0 to 100
Details
For time domain measurements, you can delay the start of a test sequence by setting a wait time.
The test sequence starts after the wait time period expires.
Example
WT 0.1
This command string sets the wait time to 100 ms.
Also see
None
IN
This command sets the time interval between sample measurements.
Usage
IN AA.AA
AA.AA
Interval in seconds: 0.01 to 10
Details
For time domain measurements, you can set the time interval between sample measurements. After
a sample measurement is made, the next measurement starts after the time interval expires.
Example
IN 0.1
This command string sets the interval to 100 ms.
Also see
None