9: Keithley External Control Interface (KXCI) Model 4200A-SCS Parameter Analyzer
Reference Manual
9-28 4200A-901-01 Rev. C / February 2017
DT
This command sets the time duration spent on each step of the sweep for a VAR1 sweep.
Usage
DT A.AAA
A.AAA
Delay time in seconds: 0 to 6.553
Details
For a VAR1 sweep, the time spent on each step of the sweep is determined by the user-set delay
time and the time it takes to make a measurement.
You typically use the delay time to allow the source to settle before making a measurement. For
example, assume a delay time of 1 s. At each step of the sweep, the source is allowed to settle for
1 s before the measurement is made.
Example
DT 1
This command string sets delay time to 1 s.
Also see
None
Measurement setup page (SM)
Use the command strings for the SM page for the following operations:
Set wait time
Set interval
Select number of readings
Select list display mode
To send the following command strings to the 4200A-SCS, select the SM page by sending the
command:
SM