EasyManua.ls Logo

Keithley 4200A-SCS

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-27
SC
This command configures the source to output a fixed voltage level for channels that are configured to be voltage
source only.
Usage
SCA, ±BBB.BBBB
A
Voltage source channel: 1 to 9; see Details
BBB.BBBB
Output voltage value: 210.00 to +210.00
Details
This command is for use with any channel configured solely as a VS1...VSn voltage source. It
configures the source to output a fixed (constant) voltage level.
If nothing is specified after the channel number, the channel is turned off.
The range of possible values for A depends on how instruments are mapped in KCon. The parameter
A represents the n in voltage source VSn.
Example
SC1, 20
This command string sets up VS1 to output a constant 20 V level.
Also see
Keithley Configuration Utility (KCon) (on page 7-1)
HT
This command sets a hold time that delays the start of a sweep.
Usage
HT AAA.A
AAA.A
Hold time in seconds: 0 to 655.3
Details
You can delay the start of a sweep by setting a hold time. When the sweep is triggered, it starts after
the hold time period expires.
Example
HT 1
This command string sets the hold time to one second.
Also see
None

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals