EasyManuals Logo

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #746 background imageLoading...
Page #746 background image
Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-35
ME
This command controls measurements.
Usage
MEA
A
Measurement action:
• SingleTrigger test, store readings in cleared buffer: 1
• RepeatTrigger test, store readings in cleared buffer: 2
• AppendTrigger test, append readings to buffer: 3
• StopAbort test:
4
Details
The ME1 or ME2 command triggers the start of the test and makes the programmed number of
measurements. The measured readings are stored in the buffer. Note that the buffer is cleared before
readings are stored.
The ME3 command also triggers the test but does not clear the buffer before storing the measured
readings. The readings are appended to the readings already stored in the buffer. The buffer can hold
up to 4096 readings.
The ME4 command aborts the test.
Example
ME1
This command string triggers the start of the test and stores the readings in the cleared buffer.
Also see
None
Data output and file commands
The command strings for the following operations are valid in any system mode page:
• Obtain output data
• Save file
• Get file

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals