9: Keithley External Control Interface (KXCI) Model 4200A-SCS Parameter Analyzer Reference Manu
al
9-34 4200A-901-01 Rev. C / February 2017
YB
This command configures the Y2-axis of the graph.
Usage
YB 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
AAAAAA
The SMU channel name for the Y2-axis, up to 6 characters; must be one of the
SMU channel names that you specify on the channel definition (DE) page
B
Y2 axis scale type:
Linear scale: 1
Log scale: 2
Log scale absolute value: 3
CCCC.CCC
Y2-axis minimum value
Volts: ±9999
Amps: ±999
DDDD.DDD
Y2-axis maximum value
Volts: ±9999
Amps: ±999
Example
YB 'I2', 2, 100E-9, 1E-3
The command string:
Specifies that values from SMU channel I2 are to be plotted on the Y2-axis.
Sets up the Y2-axis to be scaled logarithmically between 100 nA and 1 mA.
Also see
Channel definition page (DE) (on page 9-13)
Measurement control page (MD)
To send the command string to control measurements, select the measurement control page by
sending the command:
MD