EasyManua.ls Logo

Keithley 4200A-SCS

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-37
SV
This command saves a program file or data file.
Usage
SV 'A BBBBBB'
SV 'A BBBBBB CCCCCCCC'
A
File type:
Program file: P
Data/Program file: D
BBBBB
Name of file (up to 6 characters)
CCCCCCCC
Comment (up to 8 characters)
Details
You must separate the file name and the comment from each other by a space.
When saving a program file, the present instrument settings are stored in a file at the directory path:
C:\s4200\sys\KXCI
You specify the name for the file.
The get command string is used to acquire the saved file.
Example
SV 'P Setup1'
This command string saves the command sequence as a program file named
Setup1
.
Also see
GT command: Get file (on page 9-37)
GT
This command acquires (loads) the saved data file or program file.
Usage
GT 'A BBBBBB'
A
File type:
Program file: P
Data/program file: D
BBBBB
Name of file (up to 6 characters)
Details
The file type and file name must be separated by a space.
For a program file, this command launches the program. For a data file, it opens the files. When the
saved program file is recalled, the instrument returns to the settings stored in that file.
The save command string is used to save instrument settings or to store data acquired in a test.

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals