Getting Started
R&S
®
ZNA
101User Manual 1178.6462.02 ─ 29
"Converter 1" is the converter connected to VNA port 1, "Converter 2" the converter
connected to VNA port 2.
2. In the "Converter Configuration" dialog, configure the standard setup with two con-
verters:
a) For both converter ports, set "Converter Type" to "ZNA67EXT-TS"
The firmware automatically fixes "IF Input" to "Rear & Dir" (rear IF access for
high, direct access for low frequencies).
b) In the combo box above the "Splitter 1" symbol, select "Conv. LO" as LO
source.
c) For both converter ports, select "Splitter 1" as the source for the "Local" port.
Figure 3-23: Standard setup with two external test sets
3. Select "OK" to apply the converter configuration and close the dialog.
Getting started with R&S
ZNA67EXT

Table of Contents