EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 4.3.1.1 Multiport S-parameters

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Concepts and features
R&S
®
ZNA
150User Manual 1178.6462.02 ─ 29
Meaning of 2-port S-parameters
The four 2-port S-parameters can be interpreted as follows:
S
11
is the reflection coefficient of DUT port 1, i.e. the ratio between outgoing wave
b
1
and incident wave a
1
in a forward measurement with matched DUT port 2:
S
11
= b
1
/ a
1
, if |a
1
| > 0 and a
2
= 0
S
21
is the forward transmission coefficient, defined as the ratio between outgoing
wave b
2
and incident wave a
1
in a forward measurement with matched DUT port 2:
S
21
= b
2
/ a
1
, if |a
1
| > 0 and a
2
= 0
S
12
is the reverse transmission coefficient, defined as the ratio between outgoing
wave b
1
and incident wave a
2
in a forward measurement with matched DUT port 1:
S
12
= b
1
/ a
2
, if |a
2
| > 0 and a
1
= 0
S
22
is the reflection coefficient of port 2, i.e. the ratio between outgoing wave b
2
and
incident wave a
2
in a forward measurement with matched DUT port 1:
S
22
= b
2
/ a
2
, if |a
2
| > 0 and a
1
= 0
Meaning of squared amplitudes
The squared amplitudes of the incident and outgoing waves and of the matrix elements
have a simple meaning:
Table 4-4: Squared S-parameters
|a
i
|
2
Available incident power (= the power provided by a generator with
a source impedance equal to the reference impedance Z
0
) at DUT
port i=1,2
|b
i
|
2
Reflected power at DUT port i=1,2
10 log|S
ii
|
2
= 20 log|S
ii
|
Reflection loss at DUT port i=1,2
10 log|S
21
|
2
= 20 log|S
21
|
Insertion loss of forward transmission
10 log|S
12
|
2
= 20 log|S
12
|
Insertion loss of reverse transmission
4.3.1.1 Multiport S-parameters
The multiport S-parameters extend the standard 2-port S-parameters to a larger num-
ber of incident and outgoing waves. For a 4-port DUT,
a
a
a
a
SSSS
SSSS
SSSS
SSSS
b
b
b
b
4
3
2
1
44434241
34333231
24232221
14131211
4
3
2
1
where again a
i
(i = 1 to 4) denote the incident, b
i
(i = 1 to 4) denote the outgoing waves,
and the S-parameters are expressed as S
ij
(i,j = 1 to 4).
Measurement results

Table of Contents

Related product manuals