GUI reference
R&S
®
ZNA
600User Manual 1178.6462.02 ─ 29
Start Cal Sweep
Starts the calibration sweeps for the selected port and power calibration settings and
closes the dialog. The calibration is performed as described in "Calibration procedure"
on page 220.
Open the Pwr Cal Settings tab if you wish to modify the calibration procedure.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
Calibrate Only Port Frequency
For Frequency conversion measurements, by default the reference receivers are cali-
brated for all resulting frequencies – even if they are currently not relevant for some
ports. Using this switch, you can limit the calibrated frequencies to those frequencies
that are required for the related port, which results in shorter calibration times.
Power Cal dialog – Meas. Receiver
In "Meas. Receiver" mode, the "Power Cal" dialog guides you through the calibration of
a measurement receiver. To ensure accurate source signal powers, a source power
calibration must be performed before the measurement receiver calibration.
Background information
Refer to Chapter 4.5.6.2, "Measurement receiver calibration", on page 221.