Remote command: 
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Apply
This button is enabled when sufficient data have been acquired for the calibrated ports.
The button starts the calculation of the calibration data and closes the calibration wiz-
ard.
The results are applied to the related channels. The current instrument settings are
stored with the correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the Calibration Manager dialog.
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command: 
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
5.11.1.4  Calibration setup dialog
Allows you to configure the calibration setup of the available channels.
Access:
●
Channel – [Cal] > "Start Cal" > "Cal Active Channel" – "Configure/Start Calibration"
●
Calibrate All dialog > "Add + Modify"
Cal softtool