Command reference
R&S
®
ZNA
1477User Manual 1178.6462.02 ─ 29
[SENSe:]ROSCillator:SMA:OUTPut <OffOr100MHz>
Defines whether a reference clock signal shall be output to the Reference In/Out SMA
connector.
Ineffective if this connector is used as reference clock source (see [SENSe<Ch>:
]ROSCillator[:SOURce]).
Parameters:
<OffOr100MHz> OFF | F100mhz
No output or 100 MHz
*RST: OFF
Example:
*RST; :ROSC EXT; ROSC:SMA:OUTP F100
Select the Reference In BNC connector as reference clock
source and output the 100 MHz reference clock signal at the
Reference In/Out SMA connector.
Manual operation: See "Reference I/O (SMA) > Out" on page 887
[SENSe<Ch>:]ROSCillator[:SOURce] <Source>
Selects the source of the reference oscillator signal.
Parameters:
<Source> INTernal | EXTernal | ESMA
INTernal – internal 10 MHz reference oscillator
EXTernal – external reference clock via BNC connector
ESMA – external reference clock via SMA connector
The frequency of the external reference clock can be defined
using [SENSe:]ROSCillator:EXTernal:FREQuency
*RST: INTernal
Example: See [SENSe:]ROSCillator:EXTernal:FREQuency
on page 1476.
Manual operation: See "Reference Source" on page 886
7.3.14.20 [SENSe:]SEGMent...
The [SENSe:]SEGMent<Seg>... commands define all channel settings for a seg-
mented frequency sweep. A segmented sweep is activated via [SENSe<Ch>:
]SWEep:TYPE SEGMent.
The commands in this subsystem do not accept the step parameters UP and DOWN.
Numeric values can be entered directly or using the DEFault, MINimum, MAXimum
parameters.
SCPI command reference