Command reference
R&S
®
ZNA
1230User Manual 1178.6462.02 ─ 29
7.3.4 DIAGnostic commands
The DIAGnostic... commands provide access to service and diagnostic routines
used in service, maintenance and repair. In accordance with the SCPI standard, all
commands are device-specific.
Service functions are password-protected (SYSTem:PASSword[:CENable]) and are
intended for Rohde & Schwarz service staff. Refer to the service manual for more infor-
mation.
DIAGnostic:DEVice:STATe............................................................................................ 1230
DIAGnostic:DUMP:SIZE............................................................................................... 1230
DIAGnostic:PRODuct:OPTion:INFO?............................................................................. 1231
DIAGnostic:SERVice:RFPower......................................................................................1231
DIAGnostic:SERVice:SFUNction....................................................................................1231
DIAGnostic:DEVice:STATe <Filename>
Generates a system report and writes it to the specified file. See Chapter 9.3, "Collect-
ing information for technical support", on page 1768.
Setting parameters:
<Filename> String parameter containing the file name. If no path is specified,
the file is stored to the directory
C:\Users\Public\Documents\ZNA\Report; the extension
*.zip is appended automatically.
*RST: n/a
Example:
DIAG:DEV:STAT 'report_16032011_1120'
Generate a report and store it to
C:\Users\Public\Documents\ZNA\Report\
report_16032011_1120. Use the MMEMory... commands to
rename, move, or delete the file.
Usage: Setting only
Manual operation: See "Save..." on page 874
DIAGnostic:DUMP:SIZE <DumpSize>
Determines the level of detail and hence the size of the dump files created in case of
firmware exceptions.
Parameters:
<DumpSize> NONE | MINI | NORMal | LARGe | FULL
Either disables dump file creation (NONE) or determines the
level of detail.
Manual operation: See "Error Dump Type" on page 866