985User Manual 1178.6462.02 ─ 29
7.3.1.3 CALCulate:DTIMe...
Defines the properties and retrieves the results of the skew measurement provided
with the Extended time domain analysis option R&S ZNA-K20.
CALCulate<Chn>:DTIMe:DATA?..................................................................................... 985
CALCulate<Chn>:DTIMe:LIMit:FAIL?............................................................................... 985
CALCulate<Chn>:DTIMe:LIMit:FAIL:BEEP....................................................................... 986
CALCulate<Chn>:DTIMe:LIMit:LIMit................................................................................ 986
CALCulate<Chn>:DTIMe:LIMit:STATe.............................................................................. 986
CALCulate<Chn>:DTIMe:POSition...................................................................................987
CALCulate<Chn>:DTIMe:STATe...................................................................................... 987
CALCulate<Chn>:DTIMe:TARGet....................................................................................987
CALCulate<Chn>:DTIMe:DATA? [<Data>]
Queries the results of the skew measurement
Suffix:
<Chn>
.
Channel number used to identify the active trace
Query parameters:
<Data> ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the active trace's stim-
ulus axis (CALCulate<Chn>:TRANsform:TIME:XAXis). See
the table below.
Note that the skew value is positive, if, at the defined position
(CALCulate<Chn>:DTIMe:POSition), the active trace is to
the right of the reference trace.
Usage: Query only
Options: R&S ZNA-K20
Manual operation: See "Extended Info" on page 828
<Data> omitted <Data> = ALL
Stimulus axis:
time
<skew time> <skew time>,<skew distance>, <X value of ref. point on current
trace>, <Y value of ref. point on current trace>, <X value of ref.
point on reference trace>, <Y value of ref. point on reference
trace>
Stimulus axis:
distance
<skew distance> <skew distance>, <skew time>, <X value of ref. point on cur-
rent trace>, <Y value of ref. point on current trace>, <X value of
ref. point on reference trace>, <Y value of ref. point on refer-
ence trace>
CALCulate<Chn>:DTIMe:LIMit:FAIL? <Boolean>
Indicates whether the skew has passed or failed.