EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - CALCulate:FMODel:EZD...

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1033User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:DELT:TDR <Boolean>
If the Time domain analysis option R&S ZNA-K2 is installed, this command allows you
to add the TDR-based converted impedance traces to the recommended trace configu-
ration.
Parameters:
<Boolean> OFF (0)
TDR-based impedance traces added
ON (1)
TDR-based impedance traces not added
*RST: OFF
Options: R&S ZNA-K231
Manual operation: See "Trace" on page 767
CALCulate:FMODel:EZD...
Commands for Eazy Deembedding (see Chapter 4.7.15, "Eazy de-embedding based
on IEEE 370", on page 298).
CALCulate:FMODel:EZD:COUPon:MEASure..................................................................1033
CALCulate:FMODel:EZD:COUPon:MEASure:FILename.................................................. 1034
CALCulate:FMODel:EZD:COUPon:MEASure:FILename:CLEar........................................ 1034
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe]........................................................ 1034
CALCulate:FMODel:EZD:DCEXtrapolat..........................................................................1034
CALCulate:FMODel:EZD:DUT:MEASure........................................................................ 1035
CALCulate:FMODel:EZD:DUT:MEASure:FILename.........................................................1035
CALCulate:FMODel:EZD:DUT:MEASure:FILename:CLEar...............................................1035
CALCulate:FMODel:EZD<Ph_pt>:DUT[:STATe]...............................................................1035
CALCulate:FMODel:EZD:GENerate:SIDE<1|2>.............................................................. 1036
CALCulate:FMODel:EZD:IMPedance.............................................................................1036
CALCulate:FMODel:EZD:IMPedance:BWIDlimit..............................................................1036
CALCulate:FMODel:EZD:IMPedance:REFerence............................................................1037
CALCulate:FMODel:EZD:PORT:ORDer..........................................................................1037
CALCulate:FMODel:EZD:PRESet..................................................................................1037
CALCulate:FMODel:EZD:RUN:RUN.............................................................................. 1037
CALCulate:FMODel:EZD<Pt>:RUN[:STATe]....................................................................1038
CALCulate:FMODel:EZD:COUPon:MEASure
For a fixture modeling with the Easy De-embedding (EZD) tool, this command starts
the measurement of the symmetrical 2xThru coupon at the active ports (see
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe] on page 1034).
Usage:
Event
Options: R&S ZNA-K210
Manual operation: See "Measure / Measure Open / Measure Short" on page 731
SCPI command reference

Table of Contents