EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 4.5.1.1 Normalization (reflection, transmission); 4.5.1.2 Reflection OSM calibration

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Concepts and features
R&S
®
ZNA
190User Manual 1178.6462.02 ─ 29
TRM calibration.....................................................................................................194
TRL calibration......................................................................................................195
TNA calibration......................................................................................................197
Full n-Port calibration with reduced number of Through connections...................197
Complementary isolation measurement................................................................198
Mixer calibrations.................................................................................................. 199
4.5.1.1 Normalization (reflection, transmission)
A normalization is the simplest calibration type since it requires the measurement of
only one standard for each calibrated S-parameter:
One-port (reflection) S-parameters (S
11
, S
22
, ...) are calibrated with an Open or a
Short standard providing the reflection tracking error term.
Two-port (transmission) S-parameters (S
12
, S
21
, ...) are calibrated with a Through
standard providing the transmission tracking error term.
Normalization means that the measured S-parameter at each sweep point is divided by
the corresponding S-parameter of the standard. A normalization eliminates the fre-
quency-dependent attenuation and phase shift in the measurement path (reflection or
transmission tracking error). It does not compensate for directivity or mismatch errors.
This limits the accuracy of a normalization.
Manual reflection normalizations offer Complementary Match standard measure-
ments
Manual transmission normalizations support Complementary isolation measure-
ment (optional).
Complementary Match standard measurements
For reflection normalizations, the mandatory Open or Short measurements can be
complemented by optional Match measurements. Additionally, measuring a Match
standard allows you to eliminate errors due to the directivity of the internal couplers,
which improves the accuracy of reflection measurements on well-matched DUTs (high
return loss).
For reflection measurements on DUTs with low return loss, accuracy may be degraded
compared to a simple reflection normalization.
4.5.1.2 Reflection OSM calibration
A reflection OSM (full one-port) calibration requires a Short, an Open and a Match
standard to be connected to a single test port. The three standard measurements are
used to derive all three reflection error terms:
The Short and Open standards are used to derive the source match and the reflec-
tion tracking error terms.
The Match standard is used to derive the directivity error.
Calibration