GUI reference
R&S
®
ZNA
921User Manual 1178.6462.02 ─ 29
●
Among the analyzer ports driven by the same internal source, only one (per setup)
can operate as generator port. If one of them is used as independent generator
port (in the current setup), then none of them can be used as regular driving port
(in the current setup).
●
If external switch matrices are used, then only VNA ports that are not connected to
a matrix can be used as independent generators.
Independent Gen
Use the checkboxes in the "Independent Gen" column, to select the VNA ports you
want to configure as independent CW sources.
Remote command:
SOURce:INDependent<Pt>[:STATe]
SOURce:INDependent:CLO[:STATe]
Source RF Off
If "Independent Gen" is checked, you can use "Source RF Off" checkbox to switch the
corresponding independent CW source on or off.
Remote command:
SOURce:INDependent<Pt>:OFF
SOURce:INDependent:CLO:OFF
Frequency / Power
If "Independent Gen" is checked, these parameters define the generated CW sweep.
Remote command:
SOURce:INDependent<Pt>:FREQuency
SOURce:INDependent:CLO:FREQuency
SOURce:INDependent<Pt>:POWer
SOURce:INDependent:CLO:POWer
Step Att
If "Independent Gen" is checked for "VNA RF Src <Pt>" and VNA port <Pt> is equip-
ped with a source step attenuator R&S ZNAxx-B2<Pt>, this command sets the source
attenuation to be applied.
Remote command:
SOURce:INDependent<Pt>:ATTenuator
5.19.9 Generic Device tab
The "Generic Devices" tab allows you to define and configure VISA-capable external
devices that are not natively supported by the analyzer firmware.
Background information
Refer to Chapter 4.7.42, "Generic devices", on page 331.
Setup softtool