GUI reference
R&S
®
ZNA
682User Manual 1178.6462.02 ─ 29
Figure 5-34: Stimulus Dialog: Power Sweep
Except for the "Enter & Display" properties, all these parameters can also be set in the
Sweep Softtool. Depending on the channel's Sweep Type, a different set of parameters
can be modified.
Table 5-6: Parameters for frequency sweeps
Lin/Log Freq Sweep CW Mode Sweep Time Sweep
Enter & Display x x x
Start x – –
Stop x – –
Power x x x
Number of Sweep
Points
x x x
Enter & Display
The "Enter & Display" combo boxes in the "Frequency Sweep" and "Power Sweep"
sections of the "Stimulus" dialog control how frequency and power values are entered
and displayed in this dialog.
You can either take the channel perspective (f
b
, P
b
), or the perspective of a particular
source or receiver. The two can be used interchangeably to define the sweep, if the
user-defined linear transformation relating them is not constant.
Note: For frequency sweeps, source/receiver-specific linear frequency transformation
are enabled with option R&S ZNA-K4 (see Chapter 4.7.3, "Frequency conversion mea-
surements", on page 261).
5.12.2.3 Define Parallel Measurement dialog
With a 4-port R&S ZNA or a 2-port R&S ZNA with 2nd internal source, you can mea-
sure multiple DUTs (or multiple paths of a single DUT) in parallel. The "Define Parallel
Measurement" dialog allows you to configure these DUTs and to declare the connec-
Channel Config softtool