The following "Stimulus" settings are also available on the "Power" tab:
Power
Start Power / Stop Power
Start Frequency / Stop Frequency / Center Frequency / Span Frequency
Defines the sweep range for non-segmented frequency sweeps.
For a Lin Freq sweep, setting "Start Frequency" and "Stop Frequency" or "Center Fre-
quency" and "Span Frequency" are alternatives.
For a Log Freq only "Start Frequency" and "Stop Frequency" can be set.
Note: For segmented frequency sweeps the start and stop frequencies and the num-
ber of sweep points are defined per segment. See Chapter 5.10.3.2, "Define Segments
dialog", on page 553.
Remote command:
[SENSe<Ch>:]FREQuency:STARt
[SENSe<Ch>:]FREQuency:STOP
[SENSe<Ch>:]FREQuency:CENTer
[SENSe<Ch>:]FREQuency:SPAN
SYSTem:FREQuency? (query frequency range of the network analyzer)
Start Power / Stop Power
Defines the sweep range for Power sweeps.