EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 7.3.20.2 Advanced port settings

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1668User Manual 1178.6462.02 ─ 29
Parameters:
<Group Delay
Receiver Mode>
INTernal | EXTernal
[SENSe<Ch>:]FREQuency:MDELay:RPORt <receiver port number>
Suffix:
<Ch>
.
Parameters:
<receiver port
number>
Manual operation: See "IF" on page 418
[SENSe<Ch>:]FREQuency:MDELay:UTONe
<Upper tone source signal>[, <port
number>]
Selects the source for the upper tone signal that is used for the mixer delay measure-
ment.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Upper tone source
signal>
PORT | GENerator | NONE
Upper tone source:
NONE – no source selected
PORT – internal source at port <source_no>
GENerator – external generator no. <source_no>, as defined in
the System Configuration – External Generators dialog
Range: 1 to port number of the analyzer/number of external
generators
<port number> Number of the port for the internal source or of the generator (to
be omitted for source NONE)
Example: See [SENSe<Ch>:]FREQuency:MDELay:LTONe
on page 1667
Manual operation: See "Upper Tone" on page 418
7.3.20.2 Advanced port settings
Access: [Channel Config] > "Port Config" > "Advanced Port Settings".
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes...........................................................1669
[SENSe<Ch>:]FREQuency:OFFSet:PWAVes:STATe........................................................1669
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:WAVes............................................................. 1669
SCPI command reference

Table of Contents

Related product manuals