GUI reference
R&S
®
ZNA
920User Manual 1178.6462.02 ─ 29
a wave, b wave, Source
Define a physical port by assigning its reference receiver, measurement receiver and
generator, respectively.
The receivers and generators can be freely assigned, but without reusing the same
(original) physical port in different (redefined) ports.
Remote command:
[SENSe:]UDSParams<Pt>:PARam
[SENSe:]UDSParams:ACTive
5.19.8.3 Independent Generator Settings dialog
Allows you to set up one or more of the internal sources as permanent CW signals.
In contrast to the generator mode, where the internal source is still synchronized with
the related channel, in internal mode, the R&S ZNA creates a CW signal that it is inde-
pendent of all channels.
Access: System – [Setup] > "Port Setup" tab > "Independent Generator Mode..."