EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 5.11.3.2 Cal Power Config dialog

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
629User Manual 1178.6462.02 ─ 29
"Ref. Receiver after" is the default behavior.
Remote command:
SOURce:POWer:CORRection:COLLect:METHod
Embedding
Defines whether the source flatness cal uses the configured offset de-/embedding to
get the power values at the DUT (instead of the calibrated reference plane).
This setting applies to all ports and channels. Default is "Off" (no offset or de-/embed-
ding used).
Remote command:
SOURce:POWer:CORRection:COLLect:VNETworks
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the power calibration of a source port. The last configured power meter is selected
by default. See "Configured Devices" on page 893.
Remote command:
SOURce:POWer:CORRection:PMETer:ID
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see Chapter 4.7.39.1,
"Zeroing", on page 318. A message indicates that zeroing is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
5.11.3.2 Cal Power Config dialog
The "Cal Power Config" dialog adjusts the target power for the power calibration (cal
power) and defines the target power for the reference receiver calibration. These set-
tings are particularly important for test setups involving external attenuators or amplifi-
ers.
Access: Channel – [Cal] > "Pwr Cal Settings" > "Cal Power..."
Related information
The settings in the "Cal Power Config" dialog are also used to define port-specific
source powers in arbitrary mode; refer to "Arbitrary Power tab" on page 674.
Cal softtool