Command reference
R&S
®
ZNA
1575User Manual 1178.6462.02 ─ 29
Equivalent command: SOURce:POWer:CORRection:COLLect:AVERage[:COUNt].
Parameters:
<NoCalSweeps> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Max Iterations" on page 628
SOURce<Ch>:POWer:CORRection:OSOurces[:STATe]
<Boolean>
Switches off all other sources during the calibration sweep for channel <Ch>.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean> ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST: ON
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Switch Off Other Sources" on page 628
SOURce:POWer:CORRection:PMETer:ID <PowerMeter>
Selects an external power meter for the scalar source power calibration (flatness cali-
bration).
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter> Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range: 1 to number of configured external power meters
*RST: The power meter selection is not changed by a
reset of the analyzer.
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:
ACQuire] on page 1719
Manual operation: See "Power Meter" on page 629
SOURce:POWer:CORRection:PPOWer <RecCalPower>
Defines the source power the R&S ZNA uses to perform the first calibration sweep of
the source power calibration ("Reference Receiver Cal Power"). This predefined power
value is ignored if the R&S ZNA is set to use the calculated port power result
(SOURce:POWer:CORRection:PSELect CPOW).
SCPI command reference