GUI reference
R&S
®
ZNA
389User Manual 1178.6462.02 ─ 29
5.2.6.4 Intermodulation Measurements dialog
Allows you to select the measured quantities that are relevant for intermodulation mea-
surements.
Access: Trace – [Meas] > "Intermodulation" > "Intermod Product" | "Intercept Point" |
"Main Tone"
Background information
Refer to "Intermodulation quantities" on page 269.
Intermod Product tab
Allows you to select intermodulation products as measured quantities.
Access: Trace – [Meas] > "Intermodulation" > "Intermod Product"
IM<order><side>O
Absolute InterModulation product (displayed in dB units).
●
<order> defines the order of the intermodulation product. Possible values are 3, 5,
7, or 9.
●
<side> defines the position of the intermodulation product relative to the lower and
upper tones.
– Upper intermodulation products are measured at frequencies above the upper
tone.
– Lower intermodulation products are measured at frequencies below the lower
tone.
– Major denotes the lower or upper intermodulation product, whichever is larger.
The major intermodulation product reveals the worst-case performance of the
DUT.
Currently intermodulation products are always measured at DUT Output.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
IM<order><side>OR
Relative intermodulation product at DUT output.
Meas softtool