5.2.6.4 Intermodulation Measurements dialog
Allows you to select the measured quantities that are relevant for intermodulation mea-
surements.
Access: Trace – [Meas] > "Intermodulation" > "Intermod Product" | "Intercept Point" |
"Main Tone"
Background information
Refer to "Intermodulation quantities" on page 269.
Intermod Product tab
Allows you to select intermodulation products as measured quantities.
Access: Trace – [Meas] > "Intermodulation" > "Intermod Product"
IM<order><side>O
Absolute InterModulation product (displayed in dB units).
<order> defines the order of the intermodulation product. Possible values are 3, 5,
7, or 9.
<side> defines the position of the intermodulation product relative to the lower and
upper tones.
Upper intermodulation products are measured at frequencies above the upper
tone.
Lower intermodulation products are measured at frequencies below the lower
tone.
Major denotes the lower or upper intermodulation product, whichever is larger.
The major intermodulation product reveals the worst-case performance of the
DUT.
Currently intermodulation products are always measured at DUT Output.