GUI reference
R&S
®
ZNA
533User Manual 1178.6462.02 ─ 29
Remote command:
SOURce<Ch>:POWer<Pt>:CPRemeas:NREadings
SOURce<Ch>:POWer<Pt>:CPRemeas:NTOLerance
SOURce<Ch>:POWer<Pt>:CPRemeas:CFACtor
Detector (table area)
Selects the detector to be used during the calibration pre-measurement. See Chap-
ter 4.3.5.3, "Detector settings", on page 160.
Remote command:
SOURce<Ch>:POWer<Pt>:CPRemeas:DETector
Premeas Mode
Defines how the firmware performs calibration pre-measurements in the current chan-
nel:
"Continuous"
(default)
The correction is applied continuously, from one sweep to the next.
"Once"
The correction is applied once, at the initial start of the measurement.
The Max Iterations/Tolerance/Convergence (table area) settings
define the convergence logic.
Remote command:
SOURce<Ch>:POWer:CPRemeas:MODE
5.8.3 Time Domain X-Axis tab
If the active trace is a time domain trace, these settings define its stimulus axis.
Time domain analysis requires option R&S ZNA-K2. If this option is not installed, the
"Time Domain X-Axis" tab is hidden.
Related information
Refer to the following sections:
●
Chapter 4.7.2, "Time domain analysis", on page 247
●
Chapter 5.5.5, "Time Domain tab", on page 447
Stimulus softtool