Command reference
R&S
®
ZNA
1578User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Value> Positive decimal number
Increment: 0.001
Manual operation: See "Max Iterations/Tolerance/Convergence (table area)"
on page 532
SOURce<Ch>:POWer:CPRemeas:CSTate <Boolean>
Activates/deactivates cal premeas for channel <Ch>.
In addition, cal premeas can be activated/deactivated at the port level (SOURce<Ch>:
POWer<Pt>:CPRemeas[:STATe] on page 1580).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1) | OFF (0)
*RST: OFF (0)
Manual operation: See "Cal Premeas On" on page 532
SOURce<Ch>:POWer<Pt>:CPRemeas:DETector <Detector>
Selects the detector to be used during the calibration pre-measurement.
This setting is channel- and port-specific.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Detector> SAMPle | AVERage | RMS | PEAK
Manual operation: See "Detector (table area)" on page 533
SOURce<Ch>:POWer:CPRemeas:MODE <Precal mode>
Defines how the firmware perfoms calibration pre-measurements in channel <Ch>:
Suffix:
<Ch>
.
Parameters:
<Precal mode> ONCE | PSWeep