GUI reference
R&S
®
ZNA
926User Manual 1178.6462.02 ─ 29
The "Device Name", "Interface", "Address", "Model", "Serial" and "State" columns allow
you to identify the connected device and to determine its operational state.
Config. File/Command
To configure a device for the channel selected in the title bar of the dialog, first pick a
configuration file for this device (column "Config. File") and then select one of the com-
mands defined in this configuration file (column "Command").
Note:
●
The recall set only stores the paths of the command files, but not their contents.
When loading the recall set, the referenced command files must be available at the
same locations.
●
In case loading a command file fails, the Error Log provides more information.
Remote command:
[SENSe<Ch>:]GDEVice:SELect
Reset
"Reset" unconfigures the command assignment for the selected device.
Remote command:
[SENSe<Ch>:]GDEVice:DELete <Address>
Clear
"Clear" unconfigures the command assignment for all devices.
Remote command:
[SENSe<Ch>:]GDEVice:DELete
5.19.10 Inline Cal System tab
The "Inline Cal System" tab gives access to inline calibration systems connected to the
R&S ZNA via USB.
Background Information
See Chapter 4.5.5.4, "Inline calibration", on page 211
5.19.10.1 Inline Calibration Systems dialog
USB-connected Inline Calibration Controllers (ICCs) are autodetected by the analyzer
firmware.
Setup softtool