GUI reference
R&S
®
ZNA
405User Manual 1178.6462.02 ─ 29
Configure Device Adress
Opens an additional dialog that allows you to specify the addresses of external devices
the external DLL requires to complete its tasks. For each task type, a single device
address can be configured.
The address syntax is DLL-specific: the DLL manages the connection to the external
device. In the example below, the DLL expects a VISA resource string and the VNA
operator indicates that the device is connected to (virtual) COM port 4.
Remote command:
[SENSe<Ch>:]CDLL:TASK:ADDRess
5.2.12.3 Power Added Efficiency (PAE) dialog
The "Power Added Efficiency (PAE)" dialog configures the measurement of the PAE of
an active 2-port device.
PAE plugin
The PAE measurement is preinstalled on instruments shipped with FW V2.15 or
higher. For older instruments, it is available as a separate installer that can be
downloaded from the R&S ZNA firmware page, and that is supported with FW
V2.15 or higher.
The PAE measurement requires an external power supply R&S HMP2030 that is
connected to the R&S ZNA via USB.
There are no dedicated remote commands for the PAE measurements. Instead,
the measurement parameters are provided as:
trace definition parameter "Src Port" in the Trace Definition Dialog
"Additional Input" string in the Trace Definition Dialog and External DLL Config-
uration Dialog (for permanent tasks).
Access: Trace – [Meas] > "S-Params" > "External DLL" > "Power Added Efficiency"
Background information
Refer to section Chapter 4.7.44, "Power added efficiency", on page 334.