Command reference
R&S
®
ZNA
1012User Manual 1178.6462.02 ─ 29
Parameters:
<EyeView> STIMulus | EMPHasis | JITTer | NOISe | DUT | EQUalization
The rightmost building block that shall be part of the calculation
chain STIMulus > EMPHasis > JITTer > NOISe > DUT > EQUali-
zation.
Note that the selected block must be active, for example
CALCulate1:EYE:VIEW JITTER requires
CALCulate<Chn>:EYE:JITTer:STATe to be ON.
Conversely, if CALCulate1:EYE:VIEW is set to JITTER and
subsequently CALCulate1:EYE:JITTer:STATe is turned
OFF, the calculation chain is further shortened to the next active
building block.
Options:
R&S ZNA-K20
Manual operation: See "[Slider]" on page 813
7.3.1.5 CALCulate:FILTer[:GATE]...
The CALCulate:FILTer[:GATE]... commands define the properties of the time
gate which is used to optimize the time domain response.
CALCulate:FILTer[:GATE]:TIME:AOFFset....................................................................... 1012
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer...............................................................1013
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev....................................................... 1013
CALCulate<Chn>:FILTer[:GATE]:TIME:LINK................................................................... 1014
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe................................................................1014
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW................................................................ 1015
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN..................................................................1015
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt.................................................................1015
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP................................................................. 1015
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe................................................................. 1016
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]................................................................ 1016
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow............................................................. 1017
CALCulate:FILTer[:GATE]:TIME:AOFFset <Boolean>
Activates the operating mode where the time gate is moved in the opposite direction
when the "Delay" setting is changed.
Parameters:
<Boolean> ON | OFF - enable or disable "Adjust Time Gate".
*RST: OFF