GUI reference
R&S
®
ZNA
696User Manual 1178.6462.02 ─ 29
Remote command:
[SENSe<Ch>:]PULSe:PERiod
Basic pulse generator settings
If option R&S ZNA-K7 is not available, the port-related pulse generators can only oper-
ate at a fixed duty cycle.
Fixed Duty Cycle/Duty Cycle
If "Fixed Duty Cycle" is enabled, the analyzer automatically adjusts the pulse width of
the port-related pulse generators to keep the "Duty Cycle" at the specified percentage.
The same pulse width and the minimum possible pulse delay are used for all sources.
These settings are only enabled, if Master Pulse Clock Source is set to "Internal".
Remote command:
[SENSe<Ch>:]PULSe:FXDCycle[:STATe]
[SENSe<Ch>:]PULSe:DUTYcycle
Advanced pulse generator settings
Advanced settings for the port-related pulse generators are only available, if the
R&S ZNA is equipped with option R&S ZNA-K7.
They are only active if Master Pulse Clock Source is set to "Internal" and Fixed Duty
Cycle is not selected.
Coupled Settings
Couples the settings of the port pulse generators, i.e. enforces identical settings for all
ports.
Mode softtool