GUI reference
R&S
®
ZNA
819User Manual 1178.6462.02 ─ 29
Active
Enables/disables noise insertion.
Remote command:
CALCulate<Chn>:EYE:NOISe:STATe
RMS
Defines the root mean square noise level.
Remote command:
CALCulate<Chn>:EYE:NOISe:RMS
DUT
The "Define DUT" dialog allows you to switch between different transmission (and
reflection) paths of the DUT and to configure its DC properties. In addition, it allows
you to switch between the real DUT (with measured frequency response) and an ideal
one (with flat frequency response), which can be useful when configuring the other eye
diagram blocks (see Chapter 5.17.2.6, "Advanced Settings dialog", on page 812).
Access: "DUT" button in the Advanced Settings dialog
This dialog is available with Extended time domain analysis option R&S ZNA-K20 only.
Measurement / Topolgy
see "Measurement / Topology" on page 809
DC Value
See "DC Value" on page 452
Mode
Allows you to switch temporarily between the real DUT (with measured frequency
response) and an ideal one (with flat frequency response).