GUI reference
R&S
®
ZNA
908User Manual 1178.6462.02 ─ 29
Access::
System – [Setup] > "Frequency Converter" > "Leveling Dataset..."
Channel – [Cal] > "Use Cal" > "Active Power Cals..."
Background Information
See Chapter 4.7.7.1, "Leveling", on page 283.
Power Control
Select whether you want activate converter power control, if a leveling dataset is avail-
able.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:STATe
Available
Shows the availability of a leveling data set.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe?
Load/Save
Use "Save" to persist the leveling dataset of the respective port to a *.lev file
(R&S ZNA-proprietary, binary file format). Use "Load" to recall a leveling dataset from a
*.lev file.
Remote command:
MMEMory:LOAD:CORRection:SLEVeling<PhyPt>
MMEMory:STORe:CORRection:SLEVeling<PhyPt>
Remove
Deletes the (global) leveling data set of the respective port.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:DELete