Command reference
R&S
®
ZNA
1337User Manual 1178.6462.02 ─ 29
Suffix:
<Port>
.
Splitter number
Parameters:
<Local> NONE | RLO | PORT | GENerator
The port type, further specified by its index <LocalIndex>
NONE
No LO source defined
RLO
The optional LO Out port at the rear panel of the R&S ZNA
PORT
A VNA port
GENerator
An external generator port
<LocalIndex>
The LO source port index, if required.
For RLO the index is ignored.
Options: R&S ZNA-K8
Manual operation: See "Splitter <i>" on page 904
[SENSe:]CONVerter:SPLitter<Port>:LOSS <Loss>
Defines the (cable and) splitter loss for splitter no. <Port>.
Suffix:
<Port>
.
Splitter number
Parameters:
<Loss> Loss value
Default unit: dB
Options: R&S ZNA-K8
[SENSe:]CONVerter<Port>:STATe <Boolean>
Activates or deactivates converter <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
Parameters:
<Boolean>
Options: R&S ZNA-K8
Manual operation: See "Active" on page 903
7.3.14.5 SENSe<Ch>:CORRection:ADVanced
Remote control commands for advanced calibration features.
SCPI command reference