Command reference
R&S
®
ZNA
1341User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator:STATe
<Boolean>
Suffix:
<Ch>
.
<PhyPt>
Parameters:
<Boolean>
Manual operation: See "Source Step Att" on page 593
7.3.14.6
[SENSe:]CORRection:CKIT...
The [SENSe:]CORRection:CKIT... commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of Chapter 7.3.14.7, "[SENSe:]CORRection:CKIT... with Labels",
on page 1349.
[SENSe:]CORRection:CKIT:<StandardType>.................................................................. 1341
[SENSe:]CORRection:CKIT:CATalog?............................................................................ 1345
[SENSe:]CORRection:CKIT:DELete............................................................................... 1345
[SENSe:]CORRection:CKIT:DMODe.............................................................................. 1346
[SENSe:]CORRection:CKIT:LABel................................................................................. 1346
[SENSe:]CORRection:CKIT:SELect............................................................................... 1347
[SENSe:]CORRection:CKIT:<ConnType>:SELect............................................................ 1347
[SENSe:]CORRection:CKIT:STANdard:CATalog?............................................................ 1348
[SENSe:]CORRection:CKIT:STANdard:DATA?................................................................ 1348
[SENSe:]CORRection:CKIT:<StandardType> <ConnType>, <CalKitName>,
<StandardLabel>, <MinFreq>, <MaxFreq>[, <DelayParam>, <Loss>, <Z0>[,
<C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt | MATCh |
<Resistance>[, <Port1>[, <Port2>]]]]]
[SENSe:]CORRection:CKIT:<StandardType>? <ConnType>, <CalKitName>[,
<Port1>[, <Port2>]]
Defines the parameters of a (possibly non-ideal) 1 port or 2-port calibration standard
<StandardType> within a particular cal kit. Depending on the standard type, only a
subset of the parameters is used; see Table 7-13
Note
As the query does not reflect the delay mode specified using [SENSe:]CORRection:
CKIT:DMODe, it is deprecated and only available for backward compatibility reasons.
Use [SENSe:]CORRection:CKIT:STANdard:DATA?? instead.
SCPI command reference