Command reference
R&S
®
ZNA
1342User Manual 1178.6462.02 ─ 29
Suffix:
<StandardType>
.
<string>
Standard type
For one-port standards, the first character denotes the gender,
for transmission standards the first two characters denote the
genders on both ends, e.g. MOPen for a male Open standard or
FFTHrough for a Through standard with female connectors.
For a complete list of standards, refer to Table 7-15.
Parameters:
<StandardLabel> Additional string label for the standard, typically the standard's
serial number
CalStandardProps Parameters <MinFreq>, ..., OPEN | SHORt | MATCh | <Resist-
ance> define the properties of the calibration standard. See
Parameter list.
Note: Sliding Match and Attenuation standards have only 2
parameters (<MinFreq>,<MaxFreq>). Through and Line stand-
ards only have 5 parameters (<MinFreq>,...,<Z0>).
<Port1>, <Port2> Optional port restriction.
For a one-port standard, the validity of the characterization can
be restricted to a single port. For a two-port standard, it can be
restricted to a port pair (specified using ascending port num-
bers).
Note: with a port restriction, the defined standard becomes sex-
less. Hence, for each port (pair) there can be only one standard
of a given type, i.e. :SENSe:CORRection:CKIT:MOP
<ParameterList>,1 and :SENSe:CORRection:CKIT:FOP
<ParameterList>,1 overwrite each other.
Parameters for setting and query:
<ConnType>,
<CalkitName>
String parameters uniquely identifying the cal kit to which the
standard belongs.
Note: If the specified cal kit does not exist, it is created with the
specified calibration standard.
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
Manual operation: See "One Port Standards / Two Port Standards" on page 618
SCPI command reference