Command reference
R&S
®
ZNA
1313User Manual 1178.6462.02 ─ 29
The command fails unless the conditions for Touchstone file export are met; see "Con-
ditions for Touchstone file export" on page 183. Traces are created using the
CALCulate<Ch>:PARameter:SDEFine command.
Setting parameters:
<Channel> Channel number in the active recall set.
<TouchstoneFile> String parameter to specify the name and directory of the cre-
ated Touchstone file. The file extension *.snp for an n-port
Touchstone file is mandatory.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory
<Format> COMPlex | LINPhase | LOGPhase
COMPlex - complex values (real and imaginary part).
LINPhase - Linear magnitude and phase.
LOGPhase - dB-magnitude and phase.
<ModeImpedance> CIMPedance | PIMPedance
CIMPedance (default, if omitted): normalize to a common target
impedance of 50 Ω
PIMPedance: normalize to the individual port reference impe-
dances
See "Renormalization of S-parameters" on page 184.
<Port>, <Port>, ... Physical port numbers
<Port>
Example: Suppose that a full two-port calibration for ports 1 and 2 and
channel 1 has been performed, and that a DUT with two bal-
anced ports is connected. The analyzer measures an arbitrary
mixed mode S-parameter.
MMEM:STOR:TRAC:PORT 1, 'Test_CIMP.s2p',
COMPlex, CIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to
the common target impedance and store them to a two-port
Touchstone file.
MMEM:STOR:TRAC:PORT 1, 'Test_PIMP.s2p',
COMPlex, PIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to to
the individual port reference impedances and store them to a
two-port Touchstone file.
Usage:
Setting only
Manual operation: See "Save" on page 473
7.3.12 OUTPut commands
The OUTPut... commands control the characteristics of the analyzer’s output ports.
SCPI command reference