GUI reference
R&S
®
ZNA
855User Manual 1178.6462.02 ─ 29
5.19.1.1 Controls on the Setup tab
The following buttons in the "Setup" tab open related dialogs:
●
"System Config...": System Config dialog
●
"Options...": "Info" dialog > Options tab
●
"Info...": "Info" dialog > Setup tab
●
"Operators Check...": Operators Check wizard
●
"Execute Selftest...": "Info" dialog > Selftest tab
(Runs the selftest first)
●
"Service Function...": Service Function dialog
Language
Selects the language of the graphical user interface. A message box indicates that the
vector network analyzer application needs to be restarted to activate a different lan-
guage.
English is the preinstalled language. A setup file for additional languages ("Vector Net-
work Analyzer Translation Setup") is available for download from the Rohde & Schwarz
Internet site.
Remote command:
n/a
Remote Encoding
Selects the character encoding used at the remote interface. The selected encoding
applies to directory and file names, calibration kit names, calibration unit characteriza-
tions and display titles.
Currently the following encodings are supported: ANSI (default), UTF-8, Shift JIS.
Remote command:
SYSTem:COMMunicate:CODec
Setup softtool