GUI reference
R&S
®
ZNA
469User Manual 1178.6462.02 ─ 29
Import Data to New Mem
Specifies whether the loaded data overwrite an existing memory trace, if available (box
unchecked), or whether they are used to generate a new memory trace (box checked).
If the box is unchecked and the active trace is a memory trace, then this memory trace
will be overwritten. If the box is unchecked and the active trace is a data trace, then the
data trace's last created memory trace will be overwritten (or a new memory trace will
be created, in case there was previously no memory trace assigned to this data trace).
Remote command:
MMEMory:LOAD:TRACe
5.5.11.2 Export Data - <File Type> dialog
The "Export Data - <File Type>" dialog stores data or memory traces to a trace file.
Trace files are ASCII files with selectable file format.
Access:
●
Trace – [Trace Config] > "Trace Data" > "Export snp Files" – ...
●
Trace – [Trace Config] > "Trace Data" > "Export" – ...
Data export can serve many purposes, e.g.:
●
To process and evaluate measurement data in an external application.
●
To store measurement data and reimport it in a future measurement session.
Background information
Refer to the following sections:
●
Chapter 4.4.2, "Trace files", on page 176.
●
Chapter 4.4.2.3, "Finding the best file format", on page 185
Traces softtool