GUI reference
R&S
®
ZNA
394User Manual 1178.6462.02 ─ 29
5.2.7.2 Amplifier Compression setup dialog
The "Amplifier Compression" setup dialog is an instance of the Multi-channel setup dia-
log. It allows you to set up one or more channels for compression point over frequency
measurements on non-frequency converting DUTs.
Access: Meas softtool > "S-Params" > "Gain Compression" > "Amplifier Compression"
Figure 5-8: Amplifier Compression dialog
Driving Port/Receiving Port
Selects the driving and receiving port for the gain compression measurement.
Remote command:
[SENSe<Ch>:]FREQuency:COMPression:SRCPort
[SENSe<Ch>:]FREQuency:COMPression:RECeiver
Sweep Type
Compression measurement channels are set up for linear frequency sweeps. For each
point on the (linear) frequency grid, the firmware performs a power sweep.
Start Frequency/Stop Frequency
See the corresponding controls on the Stimulus tab.
Number of Points
See the corresponding controls on the Sweep Params tab.
Base Power/IF Bandwidth/Step Att Receiving Port
See the corresponding controls on the Pwr Bw Avg softtool.
Start Power/Stop Power
Defines the start and stop power for the power sweeps that are performed at each
point on the linear frequency grid defined using Start Frequency/Stop Frequency.
Meas softtool