Command reference
R&S
®
ZNA
1425User Manual 1178.6462.02 ─ 29
Example: See [SENSe<Ch>:]FREQuency:COMPression:POINt
on page 1423.
Manual operation: See "Start Power/Stop Power" on page 394
[SENSe<Ch>:]FREQuency:COMPression:SKIP <Boolean>
For two-dimensional gain compression measurements, this command activates or
deactivates skipping the linear part, as described in "Skipping the linear part"
on page 392.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Skip irrelevant power sweep points. See [SENSe<Ch>:
]FREQuency:COMPression:SKIP:OFFSet
OFF (0)
Always sweep from [SENSe<Ch>:]FREQuency:
COMPression:POWer:STARt) to [SENSe<Ch>:
]FREQuency:COMPression:POWer:STOP.
*RST: OFF
Manual operation: See "Skip Linear Sweep Section" on page 393
[SENSe<Ch>:]FREQuency:COMPression:SKIP:OFFSet <Value>
For two-dimensional gain compression measurements, this command allows you to
define the power sweep range to be skipped, if skipping the linear part is active
([SENSe<Ch>:]FREQuency:COMPression:SKIP ON).
See "Skipping the linear part" on page 392.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> For the next frequency, after measuring at the start power
([SENSe<Ch>:]FREQuency:COMPression:POWer:STARt),
continue the power sweep <Value> dB to the left of the "Com-
pression Point Power In" determined for the previous frequency.
Default unit: dB
Manual operation: See "Start Before Cmp" on page 393
[SENSe<Ch>:]FREQuency:COMPression:SRCPort <TestPort>
[SENSe<Ch>:]FREQuency:COMPression:RECeiver <TestPort>
Defines the driving and receiving port for a gain compression measurement in channel