GUI reference
R&S
®
ZNA
570User Manual 1178.6462.02 ─ 29
Access: Channel – [Cal]
5.11.1 Start Cal tab
The "Start Cal" tab provides access to all functions for automatic or manual calibration.
Calibration of the R&S ZNA is a fully guided process.
Background information
Refer to the following sections:
●
Chapter 4.5, "Calibration", on page 186
●
Chapter 4.5.5, "Automatic calibration", on page 206
●
Chapter 4.5.6, "Scalar power calibration", on page 218
●
Chapter 4.5.7, "SMARTerCal", on page 224
●
Chapter 4.1.7, "Data flow", on page 120
5.11.1.1 Measurement modes and default calibration setup
When you open the Calibration wizard, the Calibration setup dialog, or the Calibrate All
dialog for the first time, the analyzer determines the preferred calibration setup based
on:
●
the channel's (or channels') measurement mode (or modes)
●
the channel's traces
●
whether a power meter is connected
The channel's (or channels') calibration setup is either written to the recall set automat-
ically, or you are asked to select between several reasonable setups (non-"Quick Start"
actions only).
Figure 5-20: Select Calibration Properties dialog: scalar mixer measurements
For the standard mode a calibration unit is used, if connected.
Cal softtool