EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Step 2: METAS Verification

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
650User Manual 1178.6462.02 ─ 29
The imported verification kit is assigned to the selected Connector type; the name of
the ZIP file is used as the verification kit name.
The name of a characterization file inside the zip container must indicate
the standard type (offset short, mismatch, 20 dB Att, 40/50 dB Att, airline, mis-
match airline, att, attenuator, stepped thru)
and, if applicable, its gender (f, m, fm, ff, mf, mm).
Imported verification kits with uncertainties are preserved and added to the Verification
Kit selection combo-box.
Step 2: METAS Verification
Step 2 allows you to perform the verification.
By selecting a button representing a verification standard, the required measurements
are performed. The raw measurement data are handed over to METAS VNA Tools that
returns the calibrated results and uncertainties and the pass/fail states of the respec-
tive verifications. The latter are indicated on the trace fields to the right of the verifica-
tion standard buttons (red = failed).
The verification standard buttons indicate the verification states of the respective verifi-
cation standards ("V" = not verified yet, "" = passed, "X" = failed). Use the Collapse
button to toggle a condensed view focusing on these verification states.
Collapse
In collapsed state, the "METAS Verification" dock widget presents a condensed view,
focusing on the verification results for the available verification standards.
Cal softtool

Table of Contents

Related product manuals