Command reference
R&S
®
ZNA
1670User Manual 1178.6462.02 ─ 29
Example: See [SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes
7.3.20.3 Internal 2nd LO (R&S ZNA-B5)
[SENSe<Ch>:]MSMode <LO Sequence Mode>
Suffix:
<Ch>
.
Parameters:
<LO Sequence
Mode>
ALTernated | CHOPped
[SENSe<Ch>:]SLAMode <Port to LO Assignment Mode>
Defines the LO usage for channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Port to LO
Assignment Mode>
OPTimized | DEFault
OPTimized
Use both LOs, if available
DEFault
Use only one LO
*RST: DEFault
Manual operation: See "LO Usage" on page 689
7.3.20.4 Other
CALCulate<Chn>:FORMat:WQPNormal.........................................................................1670
[SENSe<Ch>:]PPORt<PPort>:BSHift.............................................................................1671
[SENSe<Ch>:]PRANge................................................................................................ 1671
SYSTem:CORRection:FMPort[:STATe]........................................................................... 1671
CALCulate<Chn>:FORMat:WQPNormal <Boolean>
Enables phase normalization for wave quantity traces.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> If set to ON, the phase of the measured wave quantity is normal-
ized to the a-wave of the selected driving port.