Command reference
R&S
®
ZNA
1131User Manual 1178.6462.02 ─ 29
● CALCulate:TRANsform:DTFault......................................................................... 1131
● CALCulate:TRANsform:TIME..............................................................................1136
● CALCulate:TRANsform:VNETworks................................................................... 1144
● CALCulate:TRANsform... (other).........................................................................1193
CALCulate:TRANsform:DTFault...
The CALCulate:TRANsform:DTFault... commands set up and control a distance
to fault measurement (with option R&S ZNA-K2).
CALCulate<Chn>:TRANsform:DTFault:CENTer...............................................................1131
CALCulate<Chn>:TRANsform:DTFault:SPAN................................................................. 1131
CALCulate<Chn>:TRANsform:DTFault:DEFine............................................................... 1131
CALCulate<Chn>:TRANsform:DTFault:DELete............................................................... 1132
CALCulate<Chn>:TRANsform:DTFault:PEAK:COUNt?.................................................... 1133
CALCulate<Chn>:TRANsform:DTFault:PEAK:DATA<FaultNo>......................................... 1133
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe....................................................... 1134
CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold................................................ 1134
CALCulate<Chn>:TRANsform:DTFault:POINts................................................................1134
CALCulate<Chn>:TRANsform:DTFault:SELect................................................................1135
CALCulate<Chn>:TRANsform:DTFault:STARt.................................................................1135
CALCulate<Chn>:TRANsform:DTFault:STOP................................................................. 1135
CALCulate<Chn>:TRANsform:DTFault:STATe.................................................................1136
CALCulate<Chn>:TRANsform:DTFault:CENTer <Center>
CALCulate<Chn>:TRANsform:DTFault:SPAN <Span>
Defines the distance window of the Distance to Fault measurement using its center
and span.
See also CALCulate<Chn>:TRANsform:DTFault:STARt and CALCulate<Chn>:
TRANsform:DTFault:STOP.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Center>/<Span> Center/span value.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:CENTer 2; SPAN 2
Defines the DtF distance window of the active trace of channel 1
via its center (2m) and span (2m).
Options: R&S ZNA-K2
Manual operation: See "Start Distance / Stop Distance" on page 832
CALCulate<Chn>:TRANsform:DTFault:DEFine <CblName>, <CblPermittivity>,
<CblAtt1>, <CblFreq1>{, <CblAtt2>, <CblFreq2>}...
Defines a new user-defined cable type for Distance to Fault (DtF) measurements.
SCPI command reference