EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1135

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1135User Manual 1178.6462.02 ─ 29
An execution error is raised, if the calculated number of points is higher than the maxi-
mum number of sweep points the firmware allows (100001). To avoid this, reduce the
frequency span Δf (see [SENSe<Ch>:]SWEep:POINts on page 1511) and/or the
stop frequency d
stop
(see CALCulate<Chn>:TRANsform:DTFault:STOP) so that
2.6 · d
stop
· Δf / (v · c
0
) ≤ 100001 ,
where v denotes the velocity factor of the measured cable (see CALCulate<Chn>:
TRANsform:DTFault:SELect).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
SENSe1:SWEep:POINts?
returns the default number of sweep points (201).
:CALCulate1:TRANsform:DTFault:STARt 0; STOP 100
widens the DtF distance window to 100m.
CALCulate1:TRANsform:DTFault:POINts
starts the "Auto Number of Points calculation". Now
SENSe1:SWEep:POINts?
returns an increased number of sweep points.
Usage: Event
Options: R&S ZNA-K2
Manual operation: See "Auto Number of Points" on page 832
CALCulate<Chn>:TRANsform:DTFault:SELect <DtfSelectCable>
Selects one of the available (predefined or user-defined) cable types for Distance to
Fault measurements.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<DtfSelectCable> Name of the cable type.
Example:
CALCulate1:TRANsform:DTFault:SELect '5088-HFLR'
Selects cable type "5088-HFLR".
Options: R&S ZNA-K2
Manual operation: See "Cable Type..." on page 833
CALCulate<Chn>:TRANsform:DTFault:STARt <Start>
CALCulate<Chn>:TRANsform:DTFault:STOP <Stop>
Defines the window of the Distance to Fault measurement using its start and stop dis-
tance.
SCPI command reference

Table of Contents

Related product manuals