5.7.4 Multiple Peak tab
"Multiple Peak" search allows you to find multiple local minima/maxima at once.
Background information
Refer to "Basic marker search functions" on page 132.
5.7.4.1 Controls on the Multiple Peak tab
Max / Min
Sets up to 10 markers to the highest maxima or lowest minima in the configured Eval
Range. If a complex trace format is active (e.g. a polar diagram), the markers are set to
the measurement points with the maximum or minimum magnitude.
The required markers are created/deleted as needed.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MMAXimum | MMINimum
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute MMAXimum |
MMINimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?
Eval Range
Opens the Evaluation Range dialog that allows you to set the domain for the multiple
peak search. A modified domain takes effect the next time Max / Min is used.