Command reference
R&S
®
ZNA
1550User Manual 1178.6462.02 ─ 29
OFF
Do not use the configured offset or the single-ended deembed-
ding network.
*RST: OFF
Manual operation: See "ALC Deembedding" on page 529
7.3.15.7 SOURce:POWer:CORRection:COLLect...
The SOURce:POWer:CORRection:COLLect... commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt] <NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command: SOURce:POWer:CORRection:NREadings.
Parameters:
<NoReadings> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire] (for
equivalent command SOURce:POWer:CORRection:
NREadings)
Manual operation: See "Flatness Cal – Max Iterations" on page 628
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance <Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance> Tolerance value
Range: 0.001 dB to 1000 dB
*RST: 1 dB
Default unit: dB
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Tolerance" on page 628
SOURce:POWer:CORRection:COLLect:CFACtor <Convergence>
Specifies the convergence factor for a source power calibration.
SCPI command reference