EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 7.3.15.7 SOURce:POWer:CORRection:COLLect...

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1550User Manual 1178.6462.02 ─ 29
OFF
Do not use the configured offset or the single-ended deembed-
ding network.
*RST: OFF
Manual operation: See "ALC Deembedding" on page 529
7.3.15.7 SOURce:POWer:CORRection:COLLect...
The SOURce:POWer:CORRection:COLLect... commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt] <NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command: SOURce:POWer:CORRection:NREadings.
Parameters:
<NoReadings> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire] (for
equivalent command SOURce:POWer:CORRection:
NREadings)
Manual operation: See "Flatness Cal – Max Iterations" on page 628
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance <Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance> Tolerance value
Range: 0.001 dB to 1000 dB
*RST: 1 dB
Default unit: dB
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Tolerance" on page 628
SOURce:POWer:CORRection:COLLect:CFACtor <Convergence>
Specifies the convergence factor for a source power calibration.
SCPI command reference