Command reference
R&S
®
ZNA
1549User Manual 1178.6462.02 ─ 29
SOURce<Ch>:POWer<Pt>:ALC:TPOWer <Value>
Defines the target power value the ALC for port <Pt> in channel <Ch> stabilizes at the
selected measurement receiver (SOURce<Ch>:POWer<Pt>:ALC:RECeiver).
Only applies if the target power isn't taken from the port configuration (SOURce<Ch>:
POWer<Pt>:ALC:TPOWer:DEFault on page 1549 OFF).
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Value> Default unit: dBm
Manual operation: See "ALC: Target Power/Target Power from Port Config"
on page 529
SOURce<Ch>:POWer<Pt>:ALC:TPOWer:DEFault
<Boolean>
Defines whether the ALC target power is taken from the port configuration or can be
set using SOURce<Ch>:POWer<Pt>:ALC:TPOWer.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: ON (1)
Manual operation: See "ALC: Target Power/Target Power from Port Config"
on page 529
SOURce<Ch>:POWer<Pt>:ALC:VNETwork <Usage>
Allows you to use the configured offset or single-ended deembedding network for ALC.
Suffix:
<Ch>
.
Channel number
<Pt> Physical port number
Parameters:
<Usage> OFFSet | DEEMbed | OFF
OFFSet
Use the configured offset (--
>[SENSe<Ch>:]CORRection:EDELay[<Pt>] commands)
DEEMbed
Use the configured single-ended deembedding (-->
CALCulate<Ch>:TRANsform:VNETworks:SENDed:
DEEMbedding[<Pt>] commands)
SCPI command reference